MMT Binospec
Overview
This file summarizes several instrument specific items for the MMTO’s Binospec spectrograph.
Wavelength Calibration
Templates were created in 2023 that cover the usable range for each of the 3 gratings; the 1000 l/mm template was extended in 2026 to support red central wavelengths. All wavelengths are vacuum. The current template coverage is:
Grating |
Template file |
Range (Å) |
|---|---|---|
270 l/mm |
|
3825 – 9216 |
600 l/mm |
|
4042 – 10047 |
1000 l/mm |
|
3757 – 7211 |
For the 1000 l/mm grating, central wavelengths from roughly 4500 Å through
6450 Å are covered with full overlap on both detector sides. Settings near
the blue end of the grating range (e.g. central wavelength ~4000 Å) extend
slightly below the template’s blue limit but should still be usable, since
full_template only requires sufficient overlap to anchor the
cross-correlation; better test data are needed to confirm this corner.
Bad pixel mask
The static bad pixel mask is loaded from FITS files derived from the IDL
pipeline calibration data (badpix_binospec.fits plus the hard-coded bad
columns and detector trap regions defined in bino_mosaic.pro). This
adds roughly 12,500 individual bad pixels per detector, along with bad
columns and detector trap region masking, replacing the small set of
hard-coded bad columns previously identified from 2019 flat and bias
observations. The mask files are distributed with the package as
static_calibs/mmt_binospec/bpm_binospec_det{1,2}.fits.gz.